{"id":694545,"date":"2024-11-06T05:36:48","date_gmt":"2024-11-06T05:36:48","guid":{"rendered":"https:\/\/pdfstandards.shop\/product\/uncategorized\/ul-61010-2-0112021\/"},"modified":"2024-11-06T05:36:48","modified_gmt":"2024-11-06T05:36:48","slug":"ul-61010-2-0112021","status":"publish","type":"product","link":"https:\/\/pdfstandards.shop\/product\/publishers\/ul\/ul-61010-2-0112021\/","title":{"rendered":"UL 61010-2-011:2021"},"content":{"rendered":"
1 Scope and object<\/b><\/p>\n
This clause of Part 1 is applicable, except as follows:<\/p>\n
Replacement:<\/p>\n
Replace the second paragraph by the following:<\/p>\n
This Part 2 of IEC 61010 specifies particular safety requirements for the following types a) to c) of electrical equipment and their accessories, wherever they are intended to be used, whenever that equipment incorporates <\/p>\n
REFRIGERATING SYSTEMS<\/p>\n
as an integral part of, or separate from, the equipment and the equipment is in direct control of the <\/p>\n
REFRIGERATING SYSTEM<\/p>\n
.<\/p>\n
This document details all the requirements when up to 150 g of <\/p>\n
FLAMMABLE REFRIGERANT<\/p>\n
are used per stage of a <\/p>\n
REFRIGERATING SYSTEM<\/p>\n
. Additional requirements beyond the current scope of this document apply if a <\/p>\n
REFRIGERANT<\/p>\n
charge of <\/p>\n
FLAMMABLE REFRIGERANT<\/p>\n
exceeds this amount.<\/p>\n
Addition:<\/p>\n
Add the following text after the last paragraph:<\/p>\n
NOTE 101 Examples for <\/p>\n
REFRIGERATING EQUIPMENT<\/p>\n
include, but are not limited to, laboratory equipment such as laboratory refrigerators, freezers, refrigerated display cabinets.<\/p>\n
It is possible that all or part of the equipment falls within the scope of one or more other Part 2 standards of IEC 61010 as well as within the scope of this standard. In that case, the requirements of those other Part 2 standards will also apply. In particular, if equipment is intended to be used as a centrifuge, the requirements of IEC 61010-2-020 apply. However, when the equipment incorporates a refrigerating system and a heating function where the combination of the two introduces additional or more severe <\/p>\n
HAZARDS<\/p>\n
than if treated separately, then it is possible that IEC 61010-2-012 is applicable instead of this Part 2-011.<\/p>\n
See further information in the flow chart () for the selection process and guidance in the Introduction.<\/p>\n","protected":false},"excerpt":{"rendered":"
Electrical Equipment for Measurement, Control, and Laboratory Use – Part 2-011: Particular Requirements for Refrigerating Equipment<\/b><\/p>\n\n\n
\n Published By<\/td>\n Publication Date<\/td>\n Number of Pages<\/td>\n<\/tr>\n \n UL<\/b><\/a><\/td>\n 2021-05-13<\/td>\n 67<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n","protected":false},"featured_media":694555,"template":"","meta":{"rank_math_lock_modified_date":false,"ep_exclude_from_search":false},"product_cat":[2680],"product_tag":[],"class_list":{"0":"post-694545","1":"product","2":"type-product","3":"status-publish","4":"has-post-thumbnail","6":"product_cat-ul","8":"first","9":"instock","10":"sold-individually","11":"shipping-taxable","12":"purchasable","13":"product-type-simple"},"_links":{"self":[{"href":"https:\/\/pdfstandards.shop\/wp-json\/wp\/v2\/product\/694545","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/pdfstandards.shop\/wp-json\/wp\/v2\/product"}],"about":[{"href":"https:\/\/pdfstandards.shop\/wp-json\/wp\/v2\/types\/product"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/pdfstandards.shop\/wp-json\/wp\/v2\/media\/694555"}],"wp:attachment":[{"href":"https:\/\/pdfstandards.shop\/wp-json\/wp\/v2\/media?parent=694545"}],"wp:term":[{"taxonomy":"product_cat","embeddable":true,"href":"https:\/\/pdfstandards.shop\/wp-json\/wp\/v2\/product_cat?post=694545"},{"taxonomy":"product_tag","embeddable":true,"href":"https:\/\/pdfstandards.shop\/wp-json\/wp\/v2\/product_tag?post=694545"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}