Shopping Cart

No products in the cart.

BS ISO 17862:2013

$167.15

Surface chemical analysis. Secondary ion mass spectrometry. Linearity of intensity scale in single ion counting time-of-flight mass analysers

Published By Publication Date Number of Pages
BSI 2013 36
Guaranteed Safe Checkout
Categories: ,

If you have any questions, feel free to reach out to our online customer service team by clicking on the bottom right corner. We’re here to assist you 24/7.
Email:[email protected]

This International Standard specifies a method for determining the maximum count rate for an acceptable limit of divergence from linearity of the intensity scale in single ion counting time-of-flight (TOF) secondary ion mass spectrometers using a test based on isotopic ratios in spectra from poly(tetrafluoroethylene) (PTFE). It also includes a method to correct for intensity nonlinearity arising from intensity lost from a microchannel plate (MCP) or scintillator and photomultiplier followed by a time-to-digital converter (TDC) detection system caused by secondary ions arriving during its dead-time. The correction can increase the intensity range for 95 % linearity by a factor of up to more than 50 so that a higher maximum count rate can be employed for those spectrometers for which the relevant correction formulae have been shown to be valid. This International Standard can also be used to confirm the validity of instruments in which the dead-time correction is already made but in which further increases can or cannot be possible.

PDF Catalog

PDF Pages PDF Title
6 Foreword
7 Introduction
9 Section sec_1
Section sec_2
Section sec_2.1
Section sec_2.2
1 Scope
2 Symbols and abbreviations
2.1 Abbreviated terms
2.2 Symbols
10 Section sec_3
3 Outline of method
12 Figure fig_1
13 Section sec_4
Section sec_4.1
Section sec_4.2
Section sec_4.3
Section sec_4.3.1
Section sec_4.3.2
4 Procedure for evaluating the intensity linearity
4.1 Obtaining the reference sample
4.2 Preparation for mounting the sample
4.3 Mounting the sample
14 Section sec_4.4
Section sec_4.4.1
Section sec_4.4.2
Section sec_4.4.2.1
Section sec_4.4.2.2
Section sec_4.4.2.3
Section sec_4.4.2.4
4.4 Operating the instrument
15 Section sec_4.4.2.5
Section sec_4.4.2.6
Section sec_4.4.3
Section sec_4.4.3.1
Section sec_4.4.3.2
Section sec_4.4.3.3
Section sec_4.4.3.4
16 Figure fig_2
Section sec_4.4.4
Section sec_4.4.5
Section sec_4.5
Section sec_4.5.1
4.5 Acquiring the data
17 Table tab_c
Figure fig_3
Section sec_4.5.2
Section sec_4.5.3
Section sec_4.5.4
18 Figure fig_4
Section sec_4.5.5
Table tab_1
Section sec_4.5.6
19 Section sec_4.5.7
Section sec_4.5.8
Section sec_4.5.9
Section sec_4.5.10
Section sec_4.5.11
Section sec_4.5.12
Section sec_4.5.13
Section sec_4.5.14
20 Section sec_4.6
Section sec_4.6.1
Section sec_4.6.2
4.6 Checking the linearity
21 Section sec_4.6.3
Section sec_4.6.4
Section sec_4.6.4.1
22 Section sec_4.6.4.2
23 Figure fig_5
Section sec_4.6.4.3
Section sec_4.6.5
Section sec_4.6.5.1
24 Figure fig_6
25 Section sec_4.6.5.2
Section sec_4.6.5.3
Section sec_4.6.5.4
Section sec_5
5 Interval for repeat measurements
26 Annex sec_A
Annex sec_A.1
Annex sec_A.2
Annex A
(normative)

Computation of raster size, ion beam current, number of frames for analysis, and counts per pulse

27 Annex sec_A.3
28 Annex sec_B
Figure fig_B.1
Annex B
(normative)

Charge compensation setting

29 Annex sec_C
Figure fig_C.1
Annex C
(normative)

Ion detector setting

30 Figure fig_C.2
31 Annex sec_D
Annex sec_D.1
Annex sec_D.2
Annex D
(informative)

Instrumental factors affecting linearity

32 Annex sec_D.3
Annex sec_D.4
33 Reference ref_1
Reference ref_2
Reference ref_3
Reference ref_4
Reference ref_5
Reference ref_6
Reference ref_7
Reference ref_8
Reference ref_9
Reference ref_10
Reference ref_11
Reference ref_12
Reference ref_13
Reference ref_14
Reference ref_15
Reference ref_16
Bibliography
BS ISO 17862:2013
$167.15